Table 1 Analysis of estimated values of CTDIw from surveys of CT practice, expressed in terms of absorbed dose to air
Examination type CTDIw (mGy)
Sample size Mean SD Min 25% Median 75% Max
Heada 102 50.0 14.6 21.0 41.9 49.6 57.8 130
Face and sinusesb 20 31.7 15.9 - 19.9 28.0 35.2 -
Vertebral traumab 20 44.1 21.5 - 29.4 39.7 68.3 -
Chesta 88 20.3 7.6 4.0 15.2 18.6 26.8 46.4
HRCT of lungb 20 31.7 14.9 - 19.4 31.0 35.0 -
Abdomena 91 25.6 8.4 6.8 18.8 24.8 32.8 46.4
Liver and spleenb 15 26.1 11.3 - 15.4 25.0 34.0 -
Pelvisa 82 26.4 9.6 6.8 18.5 26.0 33.1 55.2
Osseous pelvisb 16 24.7 17.8 - 14.8 20.0 24.6 -
Notes:
a. Estimated values from UK survey data(8)
b. Dose data from Pilot Study (33)


Table 2 Analysis of estimated values of DLP from surveys of CT practice on the basis of absorbed dose to air (mGy cm)
Examination type DLP (mGy cm)
Sample size Mean SD Min 25% Median 75% Max
Heada 102 882 332 231 673 795 1045 2087
Face and sinusesb 20 259 118 158 180 204 353 506
Vertebral traumab 20 392 214 40 254 353 455 914
Chesta 88 517 243 72 349 490 649 1304
HRCT of lungb 20 200 71 100 136 199 278 312
Abdomena 91 597 281 115 415 525 774 1874
Liver and spleenb 20 658 293 151 485 651 894 1181
Pelvisa 82 443 233 68 266 416 566 1324
Osseous pelvisb 16 514 426 43 225 465 518 1758
Notes:
a. Estimated values from UK survey data(8)
b. Dose data from Pilot Study (33)


Table 3 Proposed reference dose values for routine CT examinations on the basis of absorbed dose to air
Examination Reference dose value
CTDIw (mGy) DLP (mGy cm)
Routine heada 60 1050
Face and sinusesa 35 360
Vertebral traumab 70 460
Routine chestb 30 650
HRCT of lungb 35 280
Routine abdomenb 35 780
Liver and spleenb 35 900
Routine pelvisb 35 570
Osseous pelvisb 25 520
Notes:
a. Data relate to head phantom (PMMA, 16 cm diameter)
b. Data relate to body phantom (PMMA, 32 cm diameter)



Figure 1 Histogram of CTDIw data for routine head examinations in the UK, on the basis of absorbed dose to air



Figure 2 Histogram of DLP data for routine head examinations in the UK, on the basis of absorbed dose to air



Figure 3 Histogram of CTDIair data for routine head examinations in the UK, expressed in terms of absorbed dose to air