Chapter 2
APPENDIX I
Data are tabulated below for a selection of scanner models relating to typical values of the normalised dose quantities nCTDIair and nCTDIw, and the factor PH or B, as described in Appendix I to Chapter 1. These illustrative data may be used in the absence of measured data to provide broad estimates of the reference dose quantities for CT. Further data relating to a more comprehensive range of scanners are available in the reference database on CT dosimetry that has been published on the Internet (http://www.efomp.org).
Manufacturer | Model | Applied potential (kV) |
Focus-axis distance (mm) |
Slice thickness (mm) |
nCTDIair mGy /mAs |
PMMA Head phantom (16 cm diameter) |
PMMA Body phantom (32 cm diameter) |
||
nCTDIw mGy/mAs |
PH | nCTDIw mGy/mAs |
PB | ||||||
Siemens | AR.HP | 130 | 510 | 10 | 0.335 | 0.252 | 0.75 | 0.128 | 0.38 |
Hi Q | 133 | 700 | 10 | 0.195 | 0.161 | 0.83 | 0.093 | 0.48 | |
Plus S | 120 | 700 | 10 | 0.128 | 0.110 | 0.86 | 0.062 | 0.48 | |
137 | 700 | 8 | 0.161 | - | - | 0.082 | 0.51 | ||
GE | Pace | 120 | 525 | 10 | 0.344 | 0.200 | 0.58 | 0.094 | 0.27 |
Max 640 | 120 | 525 | 10 | 0.258 | 0.158 | 0.61 | 0.064 | 0.25 | |
9800 | 120 | 630 | 10 | 0.204 | 0.143 | 0.70 | 0.063 | 0.31 | |
Philips | LX | 120 | 606 | 10 | 0.200 | 0.160 | 0.80 | 0.081 | 0.41 |
CX/Q | 120 | 606 | 10 | 0.172 | 0.149 | 0.87 | 0.070 | 0.41 | |
SR | 120 | 606 | 10 | 0.204 | 0.152 | 0.75 | 0.082 | 0.40 | |
Picker | PQ 2000a | 130 | 640 | 10 | 0.338 | 0.287 | 0.85 | 0.150 | 0.44 |
CGR | 12000 | 130 | 750 | 10 | 0.113 | 0.086 | 0.76 | 0.087 | 0.77 |
a. | Full field (Filter 0 used) |